7615 Hydraulic Pressure Controller

The Series 7615 Hydraulic Pressure Controller provides a unique, high speed approach to high pressure calibration and testing.

 

  • Ranges from 6000 to 40,000 psia (415 to 2750 bar)
  • Precision to 0.01% of range
  • High speed pressure control
  • Control stability to 0.01% full scale
  • Cycle mode for fatigue and pressure testing
  • Labview® driver available

The Series 7615 Hydraulic Pressure Controller provides a unique, high speed approach to high pressure calibration and testing. Utilizing patented control technology, the 7615 can control to the set-point in five seconds or less for a pressure step up to 4 000 psi. A variety of pressure ranges are available to 40 000 psi (2 750 bar). The unit is compatible with virtually any non-corrosive pressure media including water, synthetic oils, Fluorinert, mineral oil, and others. The 7615 is also available with the Triple Scale / Dual Sensor option which can provide up to six ranges in one instrument.

The Series 7615 Hydraulic Pressure Controller provides a Fast control mode where the pressure setpoint is achieved in typically five seconds or less, even when controlling into a 50 cc load volume! This combination of speed, pressure range and compatibility with virtually any liquid media is unique to the 7615. The 7615 is an excellent choice for automated life cycle/fatigue testing applications, high volume transducer manufacturing such as diesel fuel rail sensors and transducers, high pressure automotive transducers such as those used in braking systems and throttle body sensors along with virtually any application that involves high pressure test or calibration using a liquid media for safety, compatibility or other concerns. The 7615 increases throughput and reduces cost through speed and automation without sacrificing performance. In addition to Fast mode, the 7615 also allows the option of operating in the Normal control mode where the contact parameters are optimized to provide the fastest route of pressure change with minimal overshoot.

 

 

 

 

Pressure Ranges

Pressure Ranges 6 000 psia (415 bar)
10 000 psia (700 bar)
15 000 psia (1035 bar)
20 000 psia (1400 bar)
30 000 psia (2070 bar)
40 000 psia (2750 bar)Tare mode allows operation in gauge mode. All above ranges can be provided with the triple scale option (see table below).The Series 7615 can be supplied in a dual sensor configuration. Select any two ranges shown above
Triple Range Option All above ranges can be supplied with the triple scale option to provide three ranges with a single sensor configuration and six ranges with the dual sensor configuration. For each selected sensor, the performance will be 0.01% of each range up to 20 000 psig (1400 bar) and 0.02% for 30 000 psig (2070 bar) and 40 000 psig (2750 bar).

Triple Range Scale Ranges psi (bar)
Full Scale              Low                    Mid                      High
6000 (415)       2000 (140)        4000(280)          6000 (415)
10,000 (700)       3000 (200)        6000 (400)          10,000 (700)
15,000 (1035)     5000 (300)        10,000 (600)       15,000 (1035)
20,000 (1400)     6000 (450)        12,000 (1000)     20,000 (1400)
30,000 (2070)     10,000 (700)     20,000 (1400)     30,000 (2070)
40,000 (2750)     12,000 (900)     25,000 (1800)     40,000 (2750)

Performance

Measure Ranges to 20 000 psia (1400 bar)

  • Precision: 0.01% of range
  • Stability: 0.01% per year
  • Total Uncertainty**: 0.015% of range per year

Ranges above 20 000 psia (1400 bar)

  • Precision: 0.02% of range
  • Stability: 0.02% per year
  • Total Uncertainty: 0.028% of range per year

Precision is defined as the combined effects olinearity, repeatability, hysteresis and temperature effects over the entire operating temperature range 0 °F to 122 °F (-18 °C to 50 °C).

Expression of accuracy (uncertainty) conforms with the recommendations of the ISO guide to the expression of uncertainty in measurement and includes RSS of precision, stability, temperature effects, and the calibration standard to 2 sigma (95%).

Control
  • Ranges to 20,000 psia (1400 bar); 0.01% maximum full scale (FS)
  • Ranges above 20,000 psia (1400 bar); 0.02% maximum FS

Low control point: When controlling from a higher pressure to a lower pressure, low control point is 400 psig (27 bar). Unit can be vented to 0 psig (0 bar).

Options

Options Triple scale option
Dual sensor
External priming vent valve
Rackmount kit

General

Pressure Supply 100 to 110 psig (6.9 to 7.5 bar) dry air @ 5 SCFM peak (150 SCLM)
Test Port Connection Autoclave F250C
Supply Pressure Connection 1/4 in NPTF
Display Vacuum fluorescent, graphical
Tare Mode Tare mode included for gauge calibrations
Calibration A calibration report providing traceability to the National Institute of Standards and Technology (NIST) is provided with each instrument. NVLAP accredited calibration to ISO 17025 is available as an option.

Recommended calibration interval is one year.

Electrical 110 to 220 V ac, 50/60 Hz
Temperature Operating: 0 °F to 122 °F (-18 °C to 50 °C)
Storage: -4 °F to 122 °F (-20 °C to 50 °C)
Humidity 5 to 95% relative humidity, non-condensing
Dimensions (h x w x d) 14 in x 17 in x 26 in (35.5 cm x 43 cm x 66 cm)
Weight 150 lb (68 kg)

 

 